"uxnmr.info" (psf600)
CONFIGURATION INFORMATION
=========================
Path
: /opt/topspin3.1/conf/instr/spect/uxnmr.info
Date
: Tue Aug 7 10:37:56
2012
Release
: TopSpin Acquisition Version 3.1-pl5
Installed in :
/opt/topspin3.1
Host
: psf600
OS
: CentOS release 5.4 (Final)
CPU
: Intel(R) Xeon(R) CPU W3505 @ 2.53GHz (2 cores at 1596 MHz)
User
: guest (Guest Account)
Description
: Avance III 600
Location :
System : Avance
III NMR spectrometer
1H-frequency : 600.13 MHz
Order Number :
Configured in: /opt/topspin3.1/conf/instr/spect
IPSO: connected to
spectrometer subnet
- TCP/IP address = 149.236.99.254
- Tctrl : 1
-
Fctrls: 4
- Gctrl1: without digital preemphasis
- Rctrl : none
-
FREDs : none
- DPP : none
DRU: AQS DRU-E Z102520/01819 ECL
06.01
- TCP/IP address = 149.236.99.89
- Firmware Version = 111024
- DRU controls AQS-Rack and HPPR/2
AQS: connected to
149.236.99.89:/dev/tty10
_Slot_ SBSB _____________________Board_____________________
Number Addr Type HW-VS FW-VS
ID ECL Name Description
-----------------------------------------------------------
2 0x10 0x42 0x4 AV R
4.4 REC-1 AQS RXAD600
Z102117/867 ECL 04.04
3 0x34 0xc9 0x1
X 1.2 REF-1 REF/2-1000 Reference Board for AQS/2
Receiver
4 0x24 0x19 0x8 AD S
2.7 SGU-1 AQS SGU/2 600
Z103081/01788 ECL 02.07 FPGA-Vs=AE
5 0x25 0x19 0x8 AD S
2.7 SGU-2 AQS SGU/2 600
Z103081/01789 ECL 02.07 FPGA-Vs=AE
6 0x26 0x19 0x8 AD S
2.7 SGU-3 AQS SGU/2 600
Z103081/01790 ECL 02.07 FPGA-Vs=AE
7 0x27 0x19 0x8 AD S
2.7 SGU-4 AQS SGU/2 600
Z103081/01791 ECL 02.07 FPGA-Vs=AE
21 0x21 0xcf 0
P 2.0 PSD-1
23 0x20 0xd1 0
B 3.2 MASTER
25 0x20 0x7 0
B
MASTER AQS Rack Master
Logical Receiver/SGU configuration:
Router: none
installed
Transmitters at the spectrometer subnet:
----------------------------------------
BLA_W1345096_0511 W1345096/0511
ECL 2:
- TCP/IP address = 149.236.99.253
- Firmware VS = 20100510
-
Amplifier =
BLAXH2H300/100/150 E 200-600MHZ INR: W1345096/0511 ECL 2
- Router
= BLMMR32 RF AMPLIFIER MODULE: W1345204/0826 ECL 0
- Controller = BLA CONTROL BOARD 6
2CH: W1522050/4807 ECL 50
BLA_W1345092_0352 W1345092/0352 ECL 1:
-
TCP/IP address = 149.236.99.252
- Firmware VS = 20100510
-
Amplifier =
BLAX300 E 6-365MHZ INR: W1345092/0352 ECL 1
- Router
= BLMMR31 RF AMPLIFIER MODULE: W1345212/1128 ECL 0
- Controller = BLA CONTROL BOARD 6
2CH: W1522050/4633 ECL 50
LTRX Z109890/00064 ECL 01.02:
- TCP/IP
address = 149.236.99.20
- Amplifier = BSMS/2 LOCK
TRANSCEIVER 600: Z109890/00064 ECL 01.02
BSMS: BSMS/2 connected to
ethernet
- TCP/IP address = 149.236.99.20
- ELCB firmware version =
20110913
- GAB current limits = 0.0/X, 0.0/Y, 10.0/Z (in A)
- SCB
channels = 40
- Shim System = BOSS3-SB
- L-TRX = BSMS/2 LOCK
TRANSCEIVER 600: Z109890/00064 ECL 01.02
- Lock: on L-TRX board
Bruker Automatic Changer: device connected to 149.236.99.254:/dev/tty04
-
Capacity: 60 holders
- Firmware: 20040805
- use BACS air = yes
- BACS sx delay = 10 s
-
Sample Rail fast changer mode = no
Eurotherm VTU: device connected to
149.236.99.254:/dev/tty05
Preamplifiers :
HPPR: - HPPR/2
preamplifier connected to 149.236.99.89:/dev/tty10
Type : HPPR/2
Controller:
Application firmware = AO
15 LEDs for tuning, 15 LEDs for matching
Module 1 : 1H LNA
PN=Z003462, SN=00713 from 20100119
Module 2 : XBB19F 2HS
PN=Z003506,
SN=00842 from 20100113
Module 3 : XBB19F 2HS
PN=Z003506, SN=00000 from 20080318
Module 4 : 2H
PN=Z003473, SN=00725 from 20091219
Frequency
generation:
- F1: for SGU
- F2: for SGU
- F3: for SGU
- F4: for
SGU
RF cable connections (detected by 'confamp')
--------------------------------------------------------------------
SGU1
NORM output -> input 1 of transmitter 1 (BLAXH2H300/100/150 E 200-600MHZ
W1345096/0511 at TCP/IP 149.236.99.253)
SGU1 AUX output -> input 2 of transmitter 2
(BLAX300 E 6-365MHZ INR W1345092/0352 at TCP/IP 149.236.99.252)
SGU2 NORM
output -> input 2 of transmitter 1 (BLAXH2H300/100/150 E 200-600MHZ
W1345096/0511 at TCP/IP 149.236.99.253)
SGU2 AUX output -> TUNE signal input of
HPPR
SGU3 NORM output -> input 1 of transmitter 2 (BLAX300 E 6-365MHZ
INR W1345092/0352 at TCP/IP 149.236.99.252)
SGU3 AUX output -> input 3 of transmitter 1
(BLAXH2H300/100/150 E 200-600MHZ W1345096/0511 at TCP/IP 149.236.99.253)
SGU4 NORM output -> input 4 of transmitter 1 (BLAXH2H300/100/150 E
200-600MHZ W1345096/0511 at TCP/IP 149.236.99.253)
SGU4 AUX output -> open
Blanking
cable connections (detected by 'confamp')
--------------------------------------------------------------------
transmitter 1 (BLAXH2H300/100/150 E 200-600MHZ W1345096/0511 at TCP/IP
149.236.99.253) amplifier 1H-100W uses blanking 2
transmitter 1
(BLAXH2H300/100/150 E 200-600MHZ W1345096/0511 at TCP/IP 149.236.99.253)
amplifier X-300W uses blanking 1
transmitter 1 (BLAXH2H300/100/150 E
200-600MHZ W1345096/0511 at TCP/IP 149.236.99.253) amplifier 2H-150W needs no
blanking
transmitter 2 (BLAX300 E 6-365MHZ INR W1345092/0352 at TCP/IP
149.236.99.252) amplifier X-300W uses blanking 3
transmitter 3 (BSMS/2 LOCK
TRANSCEIVER 600 Z109890/00064 at TCP/IP 149.236.99.20) amplifier 2H-5W needs no
blanking
last changes 11.10.2012 Peter Schmieder